High-throughput probe

The parallelization of the sample throughput of flow probes leads to a time and cost reduction of the NMR measurement as more samples can be analyzed per unit time. ​Voxalytic's high-throughput probe is unique on the market and variable in its degree of parallelization.

​​Up to 4 double resonant NMR detectors equipped each with its own Z-gradient, on-chip shim system and fluidic system can be used to perform 4 individual NMR simultaneously in a wide bore superconducting magnet. For narrow bore magnets 2 NMR detectors can be installed in one probe to parallelize workflows.

Specifications:

  • 2 to 4 individual NMR detectors

  • Double channel: 1H/19F and X (13C, 31P, etc.)

  • RF efficiency: > 8 kHz/√W

  • Z-gradient: 30 G/cm/A

  • Active detection volume: 3 µL

  • External frequency lock sensor

Compatibility:

  • All vendors

  • Narrow (up to 2 NMR detectors), wide bore and special, e.g., 60 mm

  • 300 MHz to 1.2 GHz

 

Features:

  • Integrated on-chip shim system

  • Integrated flow detection system

  • Fluidic ports at probe base or top

  • Temperature control

  • Digital tuning and matching

Options:

  • Custom NMR detector arrangement

  • 3-axis gradient*

  • Wide tuning ranges down to 183W

* Coming soon (approx. 2026)